Cooling Method |
Air-cooling type |
Cooling Capacity(W) |
5800/6200 |
Heating Capacity/Quick Heating (W) |
0.0 |
Heating Capacity/Control Heating (W) |
6000 |
Temperature Control Precision (°C) |
+-(high-precision mode) 0.10 |
Water Setting Temperature Range (°C) |
-10~80 |
Circulated Water |
Pure or distilled water (electric conductivity of 1~10µS/cm), ethylene glycol solution (strength of 30% or less) |
Use Area Temperature (°C) |
5-40 |
External Dimensions(mm) |
H1250×W440×D800 |
Volume (kg) |
135 |
Rated Voltage (V) |
Three-phase AC200 |
Consumption Current/MIN (A) |
10.5/11.2 |
Consumption Current/MAX (A) |
30.5/31.2 |
Wattage/MIN (W) |
2950/3450 |
Wattage/MAX (W) |
8950/9450 |
Power Supply Capacity (kVA) |
15 |
Ciculation Pump/Rated Output (W) |
600 |
Ciculation Pump/Rated Flow Volume (ℓ/min) |
35/40 |
Ciculation Pump/Pump Lift (m) |
18/28 |
Maximum Tank Volume (ℓ) |
38 |
Coolant |
R-407C 1200g |
Noise (dB) |
65/68 |
Safety Circuit/Compressor |
Protector temperature (135°C OFF), overcurrent relay |
Safety Circuit/Refrigeration Cycle |
Pressure switch |
Safety Circuit/Circulated Water Circuit |
Water pressure sensor, Wed deletion / Irrigation / SW float filled with water detected |
Safety Circuit/Pump |
Overcurrent relay |
Safety Circuit/Fan Motor |
Overcurrent relay |
Safety Circuit/Heater |
Thermo-Over-Temperature (90°C OFF), current fuse |
Input-Output/Alarm Output |
No-voltage contact a ( AC250V / 2A / Resistive load) |
Input-Output/Water Temperature Top-Bottom Limit Alarm |
No-voltage contact a ( AC250V / 2A / Resistive load) |
Input-Output/Water Supply Valve/Operation Output |
No-voltage contact a ( AC250V / 2A / Resistive load) |
Input-Output/External Temperature Sensor Input |
Pt100 |
Input-Output/External Control Input |
Non-voltage input |
Input-Output/Emergency Shutdown Input |
Non-voltage input |
Input-Output/RS-232C |
Half-duplex, three-wire, one-on-one |
Functions |
Circulating water pressure alarm function,
Circulating water pressure display function,
Circulating water temperature alarm function,
Heater power display function,
Pump running time display function,
Anti-freezing operation function,
Digital filter function,
Password setting function,
Auto-tuning function,
Function ON / OFF timer,
Key lock function,
Error history memory function,
Cascade control function,
External temperature input function |
Safe Performance |
Timer condition compressor, close assist pump, adjuster for fixing, double safety pan |